EasyManuals Logo

RADWAG AS series User Manual

RADWAG AS series
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
- 65 -
Means of preparing mixtures without entering data to balance memory on formation
ingredients and their mass
If a user needs a documentation from a formula making process, e.g. in a form of a
printout, then set parameter 02 Automatic printout to 1 : YES.
In such case each accepted mass of an ingredient (by pressing F key) is automatically
printed on a connected printer or computer.
Set parameter 01 Prompts (Hints) to 0 : NO.
Enter Formulation mode by pressing ENTER key
Tare mass of a vessel for formula making process
Pour an ingredient no. 1 to the vessel until reaching desired target value (follow
criteria on prepared formulation).
Press F key. Mass of the first ingredient is saved in balance memory. The
parameters visible on the display will change to: IC=1, SUM=. . .
Mass indication on the display will zero.Press F key.
Repeat the process for all the ingredients set in a formulation
After weighing the last ingredient of a prepared formulation press
0/T
key.
The formula making process is completed. The indication with total sum of
prepared formulation is locked on the display.
Press PRINT key to print (send) a report from a completed formulation to a
connected printer of computer.
Statistical calculations
Statistical calculations are carried out only on total mass of prepared formulation (mass of
each ingredient in a formulation is not included in the statistics). If a user wants to run
statistics if the formula making mode, then:
1. Go to parameter 06 Statistics
2. Erase results of previous statistical calculations
3. Set parameter 06 Statistics to YES
4. Enter the formula making mode
5. Carry out required series of measurements
6. Once again go to parameter 06 Statistics
7. Enter parameter 06 02 Results
8. In order to print statistics press PRINT key

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43