26
16. PRINTOUTS
It is possible to define adjustment report printout template and GLP printout
template. To set the printouts go to < P6.Prnt > submenu.
16.1. Adjustment Report
<P6.1.CrEP> is a group of parameters allowing you to declare variables that
are to be printed on an adjustment report printout. Each variable features
accessibility attribute: YES print, no do not print. Adjustment report
is automatically generated at the end of each adjustment process.
Variables list:
No.
Name
Description
6.1.1.
CtP
Performed adjustment type.
6.1.2.
dAt
Adjustment date.
6.1.3.
tin
Adjustment time.
6.1.4.
Idb
Serial number of the scale.
6.1.5.
CdF
Difference between mass of an adjustment weight that was measured
during last adjustment and mass of a currently measured adjustment
weight.
6.1.6.
dSh
Dashed line separating printout data and signature fields.
6.1.7.
SiG
An area for the signature of an operator carrying out the adjustment.
Printouts are generated exclusively in English.
Report example: