32
Procedure:
Enter <P3.Func> menu and select given working mode.
Go to <Snn> function.
Press key, parameter values are displayed successively one by one:
StAb
Manual printout of stable weighing result. Upon pressing key at the moment
when the result is unstable (no pictogram displayed), the program first waits for
the stability condition to be met, only then printout is carried out.
nStAb
Manual printout of each weighing result. In case of unstable indication, <?> sign is
displayed in front of the 'mass frame'. Function available for non-verified scales
exclusively.
rEPL
Automatic printout of the first stable weighing result above <Lo> threshold (to set
<Lo> threshold go to <Lo> parameter).
rEPLi
Automatic printout with time interval set in [min] (to set the interval go to <Int>
parameter).
Set the respective value, press key to confirm, next go to the home
screen.
19.2.3. Automatic Printout Time Interval
Parameter enabling you to set frequency of automatic printout. Printout interval
is set in minutes with 1 [min] accuracy within 1 [min] 1440 [min] range.
Procedure:
Enter <P3.Func> menu and select given working mode.
Enter <Int> function, window for entering time interval value is displayed.
Enter the value, press key to confirm, next go to the home screen.
19.2.4. Lo Threshold
<Lo> parameter allows you to configure the function of automatic operation.
In order to save the next measurement, before carrying it out, the mass
indication must get below the set net value of Lo threshold.
Procedure:
Enter <P3.Func> menu and select given working mode.
Enter <Lo> function, window for entering Lo threshold value is displayed.
Enter the value and press key to confirm, then continue weighing.

Table of Contents