Minimum voltage per verification unit
0.4 uV
Minimum impedance of load cell
50
Maximum impedance of load cell
1200
Load cell power supply
5 V
Connection of load cells
4 or 6 wires + shield
Weighing platforms quantity
1
Multi range option
Yes
Standard interfaces
RS232 (1), Wireless communication
Optional interfaces
RS232 (2)
29. TROUBLESHOOTING
Cause
Solution
Power supply disconnected.
Connect the power supply to the scale.
Battery discharged.
Connect the power supply to the mains,
charge the battery.
No battery (not installed or
installed incorrectly).
Check if the battery is installed correctly
(polarization).
<7.4.t1> parameter set to
value enforcing scale shut-
down after particular time
interval.
Go to <P7.Othr> menu, set <7.4.t1>
parameter to 'nonE' value.
Weighing pan loaded during
the start-up.
Unload the weighing pan. Zero indication
is displayed.
Incorrect computer port set in
parameter <5.1.1.Prt>.
Enter < P5.ducE / 5.1.PC> submenu and
set correct <5.1.1.Prt> parameter value.
Incorrect transmission
parameters for the selected
computer port.
Enter <P4.Conn> menu and set correct
transmission parameters for the selected
computer port.
Incorrect printout frequency
for continuous transmission.
Enter < P5.ducE / 5.1.PC> submenu and
set correct <5.1.3.Int> parameter value.
Incorrect printer port set in
<5.2.1.Prt> parameter.
Enter < P5.ducE / 5.2.Prtr> submenu and
set correct <5.2.1.Prt> parameter value.
Incorrect transmission
parameters for the selected
printer port.
Enter <P4.Conn> menu and set correct
transmission parameters for the selected
printer port.
No variable declared in
weighing printout project.
Enter <P6.Prnt / 6.2.GLP> submenu and
declare variables that are to be printed.