R&S ESU Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
6.318 Operating Manual 1302.6163.12 - 03
FSE [SENSe<1|2>:]AVERage:AUTO ON | OFF not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]AVERage:COUNt 0 to 32767
FSE [SENSe<1|2>:]AVERage:COUNt 0 to 32767
R&S
ESU
[SENSe<1|2>:]AVERage:TYPE VIDeo | LINear command is used to
select logarithmic or
linear averaging on the
R&S ESU;
therefore parameters are
incompatible to the FSE
FSE [SENSe<1|2>:]AVERage:TYPE MAXimum | MINimum |
SCALar
command is used to
select logarithmic or
linear averaging on the
R&S ESU;
therefore parameters are
incompatible to the FSE
R&S
ESU
FSE [SENSe<1|2>:]AVERage[:STATe<1...3>] ON | OFF
FSE [SENSe<1|2>:]BANDwidth|BWIDth:DEMod <numeric_value> not available for
R&S ESU and FSET
FSE [SENSe<1|2>:]BANDwidth|BWIDth:PLL AUTO | HIGH | MEDium
| LOW
not available for
R&S ESU
R&S
ESU
FSE [SENSe<1|2>:]BANDwidth|BWIDth:VIDeo 1Hz to 10MHz
FSET [SENSe<1|2>:]BANDwidth|BWIDth:VIDeo 1Hz to 500MHz
R&S
ESU
FSE [SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:AUTO ON | OFF
FSET [SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:
EXTernal[:STATe]
ON | OFF not available for
R&S ESU
R&S
ESU
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio 0.0001 to 1 only numeric values
available. Parameter
ranges differ between
R&S ESU and FSE
FSE [SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio 0.001 to 1000 | SINe |
PULSe | NOISe
also text parameters are
available. Parameter
ranges differ between
R&S ESU and FSE
not available for FSET
R&S
ESU
FSE [SENSe<1|2>:]BANDwidth|BWIDth[:RESolution] 10Hz to 10MHz (anal.
filter) 1Hz to 10MHz
(FFT filter)
FSE: 10Hz to 10MHz
(models 20) 1Hz to
10MHz (models 30)
FSET [SENSe<1|2>:]BANDwidth|BWIDth[:RESolution] 10 Hz to 500MHz
R&S
ESU
FSE [SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
AUTO
ON | OFF
FSE [SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
MODE
ANALog | DIGital not available for
R&S ESU and FSET
Devices Command (Sheet 30 of 47) Parameter Notes