EasyManua.ls Logo

R&S ZNLE Series - Page 1025

R&S ZNLE Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
1025User Manual 1178.5966.02 ─ 20
TRIGger<Ch>[:SEQuence]:SLOPe <Slope>
Qualifies whether the trigger event occurs on the rising or on the falling edge or on the
beginning of the high / low level periods of the external TTL trigger signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<Slope> POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
Manual operation: See "Signal Type" on page 341
TRIGger<Ch>[:SEQuence]:SOURce <TrigSource>
Selects the source of the trigger events that the analyzer uses to start a measurement
sequence.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigSource> IMMediate | EXTernal | MANual | MULTiple
IMMediate
Free run measurement (untriggered)
EXTernal
EXTernal – trigger by external signal applied to the Trigger In
connector or pin 2 of the Aux. Port (R&S ZNL with option -B5
only)l.
MANual
Trigger event generated by pressing the "Manual Trigger" soft-
key
MULTiple
Multiple trigger mode, configured by
TRIGger<Ch>[:SEQuence]:MULTiple... commands
Example:
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next mea-
surement sequence when the "Manual Trigger" button is
pressed.
Manual operation: See "FreeRun / External / Manual / Multiple Triggers"
on page 339
VNA command reference

Table of Contents

Other manuals for R&S ZNLE Series

Related product manuals