Command reference
R&S
®
ZNL/ZNLE
666User Manual 1178.5966.02 ─ 20
Manual operation: See "Limit Check" on page 445
CALCulate:LIMit:CIRCle:FAIL:ALL? [<ChannelSetup>]
Returns a 0 or 1 to indicate whether or not the circle limit check has failed for at least
one channel in the referenced channel setup. 0 represents pass and 1 represents fail
Query parameters:
<ChannelSetup> Name of a VNA channel setup. Can be omitted if the active VNA
channel setup shall be referenced.
Usage: Query only
Manual operation: See "Limit Check" on page 445
CALCulate<Chn>:LIMit:CIRCle:SOUNd[:STATe] <Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded circle limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Fail beep on or off.
*RST: OFF
Example:
CALCulate:LIMit:CIRCle:STATe ON; SOUNd ON
Switch the limit check on and activate the fail beep.
Manual operation: See "Limit Fail Beep" on page 446
CALCulate<Chn>:LIMit:CIRCle[:STATe] <Boolean>
Switches the circle limit check on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
Example:
*RST; CALCulate:LIMit:CIRCle:DATA 0, 0, 0.5
Define a circle limit line centered around the origin of the polar
diagram, assigning a radius of 0.5 U.
CALCulate:LIMit:CIRCle:STATe ON; FAIL?
Switch the limit check on and query the result.
Manual operation: See "Limit Check" on page 445
CALCulate<Chn>:LIMit:CLEar
Resets the limit check results for the limit line test.
VNA command reference