676User Manual 1178.5966.02 ─ 20
Tip: To define the response values of several limit line segments with a single com-
mand, use CALCulate<Chn>:LIMit:LOWer[:DATA] or CALCulate<Chn>:
LIMit:UPPer[:DATA].
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<Response> Response value
Range: Virtually no restriction for limit segments
*RST: The default response values of a segment that is
created by defining its stimulus values only (e.g. by
means of CALCulate<Ch>:LIMit:CONTrol[:DATA]),
are -40 dB.
Default unit: NN
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
:CALC:LIM:SEGM:AMPL:STAR 5; STOP 5; :CALC:LIM:
SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a con-
stant response value of +5 dB.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1500000000,2000000000,5,5.
Manual operation:
See "Segment List" on page 436
CALCulate<Chn>:LIMit:SEGMent:COUNt?
Returns the number of limit line segments, including enabled and disabled segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
CALC:LIM:SEGM:COUNT?
Query the number of segments. The response is 1.
Usage: Query only
Manual operation: See "Segment List" on page 436