Operating modes, applications, channel setups, and result displays
R&S
®
ZNL/ZNLE
92User Manual 1178.5966.02 ─ 20
Run Single/Run Cont and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the [Run Single] and [Run Cont] keys control the
Sequencer, not individual sweeps. [Run Single] starts the Sequencer in single mode,
while [Run Cont] starts the Sequencer in continuous mode.
The "Single Sweep" and "Continuous Sweep" softkeys control the sweep mode for the
currently selected channel setup only; the sweep mode only has an effect the next time
the Sequencer activates that channel setup, and only for a channel-defined sequence.
In this case, a channel setup in single sweep mode is swept only once by the
Sequencer. A channel setup in continuous sweep mode is swept repeatedly.
5.4.2 Sequencer settings
The "Sequencer" menu is available from the toolbar.
Sequencer State .......................................................................................................... 92
Sequencer Mode ..........................................................................................................92
Sequencer State
Activates or deactivates the Sequencer. If activated, sequential operation according to
the selected Sequencer mode is started immediately.
Remote command:
SYSTem:SEQuencer on page 572
INITiate:SEQuencer:IMMediate on page 572
INITiate:SEQuencer:ABORt on page 571
Sequencer Mode
Defines how often which measurements are performed. The currently selected mode
softkey is highlighted blue. During an active Sequencer process, the selected mode
softkey is highlighted orange.
"Single Sequence"
Each measurement is performed once, until all measurements in all
active channel setups have been performed.
"Continuous Sequence"
The measurements in each active channel setup are performed one
after the other, repeatedly, in the same order, until sequential opera-
tion is stopped.
This is the default Sequencer mode.
Remote command:
INITiate:SEQuencer:MODE on page 572
5.4.3 How to set up the sequencer
In order to perform the configured measurements consecutively, a Sequencer function
is provided.
Running a sequence of measurements