Command reference
R&S
®
ZNL/ZNLE
571User Manual 1178.5966.02 ─ 20
<ChannelName2> String containing the new channel setup name.
Note that you cannot assign an existing channel setup name to
a new channel setup. If you do, an error occurs.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Usage: Setting only
INSTrument[:SELect] <ChannelType> | <ChannelName>
This command activates a new channel setup with the defined channel setup type, or
selects an existing channel setup with the specified name.
Also see
INSTrument:CREate[:NEW] on page 569
Parameters:
<ChannelType> Channel type of the new channel setup.
For a list of available channel setup types see INSTrument:
LIST? on page 570.
<ChannelName> String containing the name of the channel setup.
Example:
INST IQ
INST 'MyIQSpectrum'
Selects the channel setup named 'MyIQSpectrum' (for example
before executing further commands for that channel setup).
Manual operation: See "VNA" on page 85
11.3.2.2 Performing a sequence of measurements
The following commands control the sequencer.
For details on the Sequencer see Chapter 5.4, "Running a sequence of measure-
ments", on page 89.
INITiate:SEQuencer:ABORt............................................................................................ 571
INITiate:SEQuencer:IMMediate....................................................................................... 572
INITiate:SEQuencer:MODE.............................................................................................572
SYSTem:SEQuencer......................................................................................................572
INITiate:SEQuencer:ABORt
This command stops the currently active sequence of measurements.
You can start a new sequence any time using INITiate:SEQuencer:IMMediate
on page 572.
Usage:
Event
Manual operation: See " Sequencer State " on page 92