Command reference
R&S
®
ZNL/ZNLE
1038User Manual 1178.5966.02 ─ 20
Example:
CORR:CKIT:N50:FOPEN 'ZV-Z21','',
0,1.8E+010,0.0151,0,0,0.22,-0.22,0.0022
Define the properties of the open (f) standard for the N 50 Ω
connector type contained in the ZV-Z21 calibration kit: Assign a
valid frequency range of 0 Hz to 18 GHz, an electrical length of
15.1 mm, 0 dB loss and define the polynomial coefficients of the
fringing capacitance as 0 fF, 0.22 fF/GHz, –0.22 fF/(GHz)
2
,
0.0022 fF/(GHz)
3
.
[SENSe:]CORRection:CKIT:INSTall <CalKitFile>
Loads cal kit data from a specified R&S ZVR cal kit file.
Setting parameters:
<CalKitFile> String parameter to specify the name and directory of the cal kit
file to be loaded.
Note: The loaded file must be a R&S ZVR-specific cal kit file
with the extension *.ck. VNA cal kit files (*.calkit) can be
imported using the MMEMory:LOAD:CKIT command. Keysight
cal kit files can be imported manually and converted into
*.calkit files.
Example:
CORR:CKIT:INST 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNL\Calibration\Kits\ZCAN.ck'
Load the previously created R&S ZVR cal kit file ZCAN.ck from
the default cal kit directory.
MMEM:STOR:CKIT 'ZCAN', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNL\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a VNA cal kit file
ZCAN.calkit (assuming that the cal kit name stored in
ZCAN.ck reads "ZCAN").
Usage: Setting only
[SENSe<Ch>:]CORRection:COLLect[:ACQuire] <Standard>[, <Dispersion>[,
<Delay>]]
Starts a calibration measurement in order to acquire measurement data for the
selected standards. The standards are reflection or transmission standards and must
be connected to port 1 or 2 of the analyzer.
Tip: Use the generalized command [SENSe<Ch>:]CORRection:COLLect[:
ACQuire]:SELected to obtain measurement data at arbitrary analyzer ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
VNA command reference