Command reference
R&S
®
ZNL/ZNLE
1047User Manual 1178.5966.02 ─ 20
POWer:S<Ptout><Ptin>
S-parameter with output and input port number of
the DUT, e.g. S11, S
21
.
POWer:RATio A<Ptout> | B<Ptin>,
A<Ptout> | B<Ptin>
Output: A<Ptout>/B<Ptin>
Ratio, e.g. B2, A1 for b
2
/a
1
drive Port 1
POWer:A<Ptout>
Wave quantity with stimulus port number of the ana-
lyzer, e.g. a
1
.
POWer:B<Ptin>
Wave quantity with receive port number of the ana-
lyzer, e.g. b
2
.
POWer:Z<Ptout><Ptin>
Matched-circuit impedance (converted Z-parameter)
with output and input port number of the DUT, e.g.
Z
11
, Z
21
.
POWer:Y<Ptout><Ptin>
Matched-circuit admittance (converted Y-parameter)
with output and input port number of the DUT, e.g.
Y
11
, Y
21
.
POWer:KFACtor
POWer:MUFactor<Lev>
Stability factor K
Stability factors μ
1
or μ
2
VOLTage[:DC] DCIN1 | DCIN2
Output: DC 1 V, DC 10 V
DC Input 1 or 2
Note: The mnemonics POWer: and VOLTage: are not used in output strings.
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire] <Sensor>
Initiates a source power calibration for the source port <PhyPt> using an external
power meter no. 1 or 2. To initiate a source power calibration for arbitrary power
meters, use the alternative commands listed in the program example below.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Setting parameters:
<Sensor> ASENsor | BSENsor
ASENsor external power meter Pmtr 1.
BSENsor external power meter Pmtr 2.
*RST: ASENsor
VNA command reference