VNA GUI reference
R&S
®
ZNL/ZNLE
473User Manual 1178.5966.02 ─ 20
Start Auto Cal
Performs a fully automatic system error calibration for the active channel, using the
active calibration unit with its factory calibration and auto-detection of ports.
If one port is active, an OSM calibration is performed. If two ports are active, an UOSM
calibration is performed.
Calibration – Start... (Cal Unit)
Opens the Calibration Unit wizard.
Calibration – Start... (Manual)
Opens the Calibration Setting wizard.
Calibration – Repeat...
Re-enters the data acquisition step of the current calibration, restoring the related
channel settings. This provides a convenient way to repeat or correct an existing cali-
bration without necessarily repeating all measurements.
Scalar Power Cal – Power Cal...
Opens the Power Cal wizard.
If the active channel setup contains several channels, an "Info" box lets you decide
whether the R&S ZNL/ZNLE acquires calibration data for the active channel or for all
channels.
Cal Validation ...
Opens the Cal Validation dock widget that allows you to set up and run a Calibration
validation.
10.12.1.2  Channel Selection for Calibration dialog
Use
●
"Active Channel" if you only want to calibrate the active channel
●
"All Channels" if you want to calibrate all channels in the current channel setup
●
"Selected Channels" if you want to calibrate only certain channels
Cal softtool