Command reference
R&S
®
ZNL/ZNLE
732User Manual 1178.5966.02 ─ 20
Manual operation:  See "Defined Value" on page 381
CALCulate<Chn>:STATistics:NLINear:COMP:RMARker <Marker>
Allows you to select the marker whose value shall be used as the reference ("small sig-
nal value") for the compression point calculation.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to MARKer.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Marker> Marker number
If marker <Marker> is OFF for the related trace, the set com-
mand is silently ignored.
Manual operation:  See "Selected Marker" on page 380
CALCulate<Chn>:STATistics:NLINear:COMP:RPHase <Phase>
Manually defines the reference phase for compression point calculation on phase for-
matted traces.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to DVALue.
For dB formatted traces, the reference level can be set using CALCulate<Chn>:
STATistics:NLINear:COMP:RLEVel.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Phase> Reference phase
*RST:  1°
Default unit: deg
Manual operation:  See "Defined Value" on page 381
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe] <Boolean>
Displays or hides the compression point result in the diagram area channel <Chn>'s
active trace..
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - statistical info field on or off.
*RST:  OFF
VNA command reference