Command reference
R&S
®
ZNL/ZNLE
791User Manual 1178.5966.02 ─ 20
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
<Cmp> Number of capacitance in circuit model. The total number of
capacitances depends on the selected circuit model.
Parameters:
<Capacitance> Capacitance C<Cmp> for the specified circuit model.
Range: -1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST: 1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel> CSL | LSC | CSC | SLC | SCL | SCC | SHLC | GSL | LSG |
GSG | SLG | SGL | SGG
Possible circuit models (character data); see Table 11-9.
Example:
*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C cir-
cuit model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:EMB:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation: See "Network" on page 551
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA <Interchange>, <arg1>
Defines a dembedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition to select the adequate circuit model before executing this com-
mand.
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt> to load circuit
data from a Touchstone file located at the R&S ZNL/ZNLE's file system.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Setting parameters:
<Interchange> FPORts | IPORts | SGATes
VNA command reference