Command reference
R&S
®
ZNL/ZNLE
821User Manual 1178.5966.02 ─ 20
DISPlay[:WINDow<Wnd>][:STATe] <Boolean>
Creates or deletes a diagram area, identified by its area number <Wnd>.
Suffix:
<Wnd>
.
Number of the diagram area to be created or deleted.
Parameters:
<Boolean> ON | OFF - creates or deletes diagram area no. <Wnd>.
*RST: -
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
Manual operation: See "Add Trace + Diagram" on page 423
DISPlay[:WINDow<Wnd>]:TITLe:DATA <Title>
Defines a title for diagram area <Wnd>.
Suffix:
<Wnd>
.
Number of the diagram area.
Parameters:
<Title> String variable for the title. The length of the title is practically
unlimited but should be kept short enough to be displayed in the
diagrams.
Example:
*RST; :DISP:WIND:TITL:DATA 'S21 Test Diagram'
Define a title for the default diagram area. The title is displayed
below the top of the diagram area.
DISP:WIND:TITL OFF; TITL:DATA?
Hide the title. The title is no longer displayed but still defined so
it can be displayed again.
Manual operation: See "Title" on page 423
DISPlay[:WINDow<Wnd>]:TITLe[:STATe] <Boolean>
Displays or hides the title for area number <Wnd>, defined by means of
DISPlay:WINDow<Wnd>:TITLe:DATA.
Suffix:
<Wnd>
.
Number of the diagram area.
VNA command reference