EasyManua.ls Logo

R&S ZNLE Series - Page 908

R&S ZNLE Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
908User Manual 1178.5966.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE <CalType>[, <Characterization>[,
<TestPort1>[, <TestPort2>]...]]
Selects and initiates an automatic calibration at arbitrary analyzer and cal unit ports.
This command also selects the calibration type. A progress monitor for the calibration
sweeps is displayed.
Tip: If the test setup contains a high attenuation the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE.
If several calibration units are connected, use SYSTem:COMMunicate:RDEVice:
AKAL:ADDRess to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort
See [SENSe<Ch>:]CORRection:COLLect:AUTO:
CONFigure on page 903.
<Characterization> Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
The empty string (' ') refers to the factory calibration of (and
stored on) the active calibration unit.
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied.
<TestPort2> Second test port number of the analyzer
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage: Setting only
Manual operation: See "Calibration Type / Source" on page 477
VNA command reference

Table of Contents

Other manuals for R&S ZNLE Series

Related product manuals