DIGITAL INCUBATOR
4. Function Settings
18
(2) Temperature Setting&Humidity Setting /
How to Set Abnormal High Temp. Alarm
After setting required value, press button to save.
Then, it will go to the next stage, abnormal low temp.
setting with display "LO" for about 0.5 sec.
Adjust abnormal high temperature value with
or button. [Default setting : 2 (2)]
How to Set Abnormal High Temperature Alarm
Abnormal High temperature:
This function gives notice when incubator temperature is higher than setting temperature because of abnor
mal ambient temperature increase.
When room temperature is higher than incubator tem
perature, this function works with buzzer. Here, "HI"
and temp. gap are displayed alternatively. Press OK
button to remove buzzer.
Step 7
Step 8
It shows that the set temperature value is approximately 3 (3)
higher than the Maru incubator’s incubation chamber(the Cabinet).
ex :
Just press button if you do not need to change setting.
* Max. abnormal temp. alarm limit (0.0 ~ 5.0 Default Setting:2.0)
From the humidity setting, press the button once
and “HI” (abnormal HIgh temperature alarm) will be
displayed for 0.5 second and temperature display will
blink.
From the Abnormal HIgh Temperature Alarm setting, press
the button once and “LO” (abnormal LOw temp.
alarm) will be displayed for 0.5 second and the temperature
display will blink.
After setting required value, press button to save.
Then, it will go to the next stage,
& change.
Abnormal Low Temperature: " LO" will indicated on LCD.
(The others, same with abnormal high temp. alarm function.)
Adjust abnormal low temp. setting with or
button. [Default Value: -3
℃ (-3℉)]
ex :
It shows that the set temperature is 2
(2) lower than the
actual cabinet temperature of the machine.
Just press button if you do not need to change setting.
How to Set Abnormal Low Temperature Alarm
This function gives notice when incubator temperature
is lower than setting temperature because of abnormal
ambient temperature decrease.
Step 9
Step 10
* Min. abnormal temp. alarm limit (0.0 ~ -5.0 Default Setting:-3.0)