CALIBRATION
26
“GrAd” xxxx (no units).
Compensation Curve Factor: This gives the correction for the change in lift
speed. If a faster lift gives a lower weight reading, then decrease the “GrAd”
figure.
“ProP” xxxx (%)
Proportional Correction Factor: This gives the reduction in speed
compensation for an empty bucket, compared to a full bucket. Sometimes
after an Autocal, it is found that the speed correction with a full bucket is OK,
but with an empty bucket there is a larger error (e.g. 50 –60kg).
If with an empty bucket, the weight reading is lower at a slower lift speed than
at a fast speed, then reduce the factor by 2 – 3% until the error is removed.
“FrEq” xxxx (Hertz)
Frequency at full load (determined from Autocal):
6.3.3 Live Static (Channel 2)
This sets how the weight reading is displayed in Static Weighing mode.
“OFF” The weight reading is “frozen” on the display after the weight has
been sampled.
“ON” You get a “live” weight reading while the bucket is at the weighing
position.
“AUTO” You get a live reading, but it will only change if there is a change of
5% or more in the calculated weight (default setting).
Press to toggle.