Contact Time
Clorox® Germicidal Wipes 1 minute
Note
Your meter and lancing device should be CLEANED and DISINFECTED
a minimum of once per week and immediately when blood is visible,
for up to 2 cleaning and disinfection cycles per week over the 4-year
use life of the device.
Over time, cleaning and disinfecting may cause the meter and lancing
device to deteriorate. This system has been validated to a 4-year use life
by exposing the meter and lancing device to 832 wipes (416 cycles) with
the disinfectant product listed on page 37 of this user manual. This
supports cleaning and disinfection twice per week over the 4-year use
life of the device.
DISINFECTING
Two wipes must be used; one wipe to clean and a second wipe to disinfect.
After cleaning your meter or lancing device, repeat steps 1, 2 and 3 using a
second germicidal wipe to disinfect.
Allow the meter and lancing device to remain visibly wet for the full contact
time detailed below.