Possibility of acoustic signals not being heard
Loud grinding noises
- Acoustic alarms and voice communication might not be heard.
• Consider the volume of the grinding noise in relation to other
acoustic signals in the work environment. You may wish to use
additional visual signals.
4.7 Noise levels RS200:
Noise measurement in accordance with DIN 45635-31-01-KL3
The noise levels are largely influenced by the machine speed, the grinding material and the
grinding set.
Workplace-related emissions value LpAeq = up to 84dB(A)
Sound power level LWA = 99dB(A)
Measurement conditions:
Grinding set: 250ml steel with ring (90/125) and Puk (65)
Sample material: 100g cement clinker, particle size <2mm, 6 C20 grinding aid tablets
Speed: 1450min
-1
Sound level meter : Brüel & Kjaer 2237 Controller
4.8 Degree of Protection
IP40
4.9 Protective Equipment
The device is equipped with automatic lid closing which prevents it being started in an unsafe
state.
– The device can only be started with closed lid .
– The lid can only be opened when the device has come to a halt.
– The device can only be started with taut clamping lever.
– The device is switched off automatically if the clamping lever comes loose
during grinding.
4.10 Dimensions and Weight
Height: up to approx. 1220mm / Width : 820mm / Depth: up to approx. 780mm
Weight : RS200 net approx. 210kg
4.11 Required Floor Space
Height (open hood): 1900mm / Width: 820mm /
Depth: 780mm;
A clearance distance of 100mm is necessary at the back to guarantee space to operate the
main switch.