Cleaning, Wear and Maintenance
41
7 Cleaning, Wear and Maintenance
Risk of a fatal electric shock
- An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
• Do not clean the blender under running water. Use only a cloth dampened
with water.
• Disconnect the power supply plug before cleaning the blender.
W0012
The device must always be switched off and disconnected from the mains before any
interventions for cleaning or servicing purposes.
7.1 Cleaning
: Closing pin maintenance
The following maintenance work should be carried out at regular intervals, and at least monthly
to guarantee the operational safety of the device:
• Check roller (1) of the closing pin for free movement, and oil where
necessary, e.g. using sewing machine oil.
• Clean the magnets (2) on the closing pin.
7.2 Replacing the machine fuses
Mortal danger from electric shock
Exposed power contacts
– When replacing fuses on the cutout or fuse adapter you may come into
contact with live contacts. An electric shock can lead to burns and to
cardiac arrhythmias or to respiratory arrest and cardiac arrest.
• Remove the mains cable before replacing fuses.