: Changing the fuses
NOTICE
Always replace both fuses (TB).
– Type of fuse: 2 x 200mA T 250V
• Unscrew the fuse holders (TC1) and (TC2).
• Remove the fuse from the fuse holders and insert the new fuses.
• Place the fuse holders with the inserted fuses in the openings (TA).
7.3 Maintenance
Risk of loss of life through electric shock
Strong voltage through capacitor discharge
– Due to capacitor discharge on the frequency inverter, the
device conducts voltage for up to 3 minutes after the plug
has been pulled out.
– You may come into contact with live contacts when the
device is open. An electric shock can lead to burns and
arrhythmia or to respiratory failure and heart failure.
• After removing the mains lead, wait 3 minutes before
opening the device.