4.2 Grinding jar nominal volume
Wear or damage to the grinding set
No filling or insufficient filling quantity
– Increased wear or damage to the grinding set is possible if the grinding set
is operated without filling or with insufficient filling qantity.
• The grinding set must be filled to at least 1/3 of the nominal volume.
4.3 Feed Grain Size
In addition to the instrument settings, the filling level of the grinding jar is also of crucial
importance for a successful grinding process in the Vibratory Disc Mill of Retsch GmbH. The
table indicates approximate values for the recommended amount of sample as a function of the
grinding set used. A median filling (sample quantity) results in best grinding efficiency and
increases the life time of the grinding sets.
Grinding set
nominal
volume [ml]
4.4 Rated Power
1500W
Make sure that the voltage and frequency of your mains connection corresponds to that on the
type plate of the device. The mains connection must be fused to at least 16A.
Wear or damage of the device
Operation without grinding set
− During operation of the device without grinding set, excessive wear or
damage to the device may occur.
• Operate the device only with a grinding set mounted.
4.5 Motor rotation speed
The speed can be adjusted within the following range:
– 700 - 1500min
-1
In the case of agate grinding sets, it is not possible to set a speed greater than 700min
-1
.