If the LED light is solid green, there is good contact and the
ultrasonic energy is being emitted.
If the treatment is paused or the sound head is not connected
properly, there will be no green LED light on the top of the sound head.
CAUTION:
The movement performed with the sound head should be applied with
regular movement, not too slow to avoid inducing heat, nor too fast to prevent
a bad contact which would reduce the eectiveness of the treatment.
WARNING:
If replacing the sound head is necessary, the device must be switch o
before disconnecting and replacing the sound head.
5.5.5.5 After the Treatment
Clean the treatment area well before treatment as well as the
treatment head by using a towel or a tissue.
The treatment head should be cleaned with a 70% alcohol
solution.
Check if there are any signs of improvement (e.g. pain,
circulation and mobility).
When entering the next treatment session, the patient is
instructed to report any possible reaction.
5.5.5.6 Pause or Stop Treatment
Press the central controller dial to pause the treatment. Press the dial
again to continue the treatment.
Emergency stop
Press
to stop treatment immediately.