27
16. To Pause treatment, first press CH2 knob
and then press the central controller dial to
pause the treatment.
5.4 The Default Parameters
Each therapeutic waveform has 10 programs, you can set and save
the parameters of all programs, the details about default parameters
please refer to below:
Waveform Pro-
gram
Phase CC/
CV
Vector
(auto)
Vector
(manual)
(°)
C.F.
(kHz)
Beat
H.
(Hz)
Beat
L.
(Hz)
Ultra-
sound
(MHz,
50%)
Treat.
Time
(min)
Interferential
Traditional
(4 pole)
1 1 CC 0 45 4.0 110 100 1 15
2 CC 0 45 4.0 110 100 1 0
3 CC 0 45 4.0 110 100 1 0
2 1 CC 0 45 4.0 150 100 1 10
2 CC 0 45 4.0 150 100 1 0
3 CC 0 45 4.0 150 100 1 0