33
5.5 Each Stimulation Set-Up Procedure
WARNING: When using IF-4P, you must use 2 channels
(4 electrodes) and criss cross the channels as shown in the
picture to the right. If using IF-2P, only use one channel
(two electrodes).
5.5.1 4-Pole Interferential Stimulation Set-up Procedure
1. In order to turn on the device, press ON/
OFF switch to [
] icon which is located on
the side of the device
2. When you turn the ComboCare™ on, the
device will perform a self-check for about
6~8 seconds, and then the default parameters
will display the last treatment mode.
3
. Press channel 1 or 2 button to enter channel
1 and channel 2 parameter setting mode.
4. Press the [Waveform] button to choose
the therapeutic waveform, and then rotate
the central controller dial (
) to select “IF-
4P” waveform.
5. Press the [Program/Save] button to
choose between P1 – P10. Then rotate the
central controller dial (
) to until the desired
program number is displayed.
6. There are two modes in the device:
Normal/Default Mode
Professional Mode
To switch from Normal to Professional, press
and hold the [Step] button until you hear
a “beep”. The program will change from P1
to S1 for professional mode. In Professional
Mode, each program has 2 treatment steps.
Press the [Step] button again and the
display will show like the figure to the left.
The parameters of each step can be set
according to the following methods.
Program
Save
Step