5.3 SURVEY OF THE METHODS
The calculation procedures, on which all methods are traceable from the list of methods, are mentioned in the
following table. Criterion is the characteristic of the calculation procedure (see below). For detailed description of
the respectively accompanying procedure of method see chapter 5.4 - DESCRIPTION OF METHOD
PROCEDURES.
CP-No. Characteristic Method Calculation formula
CP 1 C/F Endpoint with Factor
C = F A
S
CP 2 C/F/Rb Endpoint with Factor
C = F (A
S
- A
RB
)
CP 3 C/F/Sb Endpoint with Factor
C = F |A
S
- A
SB
|
CP 4 C/F/SbRb Endpoint with Factor
C = F (|A
S
- A
SB
| - |A
RB
- A
RBB
|)
CP 5 C/S Endpoint with Standard
C = F A
S
CP 6 C/S/Rb Endpoint with Standard
C = F (A
S
- A
RB
)
CP 7 C/S/Sb Endpoint with Standard
C = F |A
S
- A
SB
|
CP 8 C/S/SbRb Endpoint with Standard
C = F (|A
S
- A
SB
| - |A
RB
- A
RBB
|)
CP 9 FTK/F/Rb Fixed Time Kinetic with Factor
C = F ( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
CP 10 FTK/S/Rb Fixed Time Kinetic with Standard
C = F ( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
CP 11 KIN/F/Rb Kinetic with Factor
C = F ( A
S,Minit
- A
RB,Minit
)
CP 12 KIN/S/Rb Kinetic with Standard
C = F ( A
S,Minit
- A
RB,Minit
)
CP 13 TRANSM. Transmission in %
CP 14 C/F DELTA Endpoint with Factor
C = F ( A
S2-Sb2
- A
S1-Sb1
)
CP 15 C/F 3 WL Measurement with 3 Wavelengths C = 168 * A
415nm
– 84 * A
380nm
– 84 * A
450nm
Explanations:
CP-No. ........................Number of the calculation procedure (chapter 6 - METHOD EDITOR)
Characteristic ..............Name of the calculation procedure (chapter 12.1 - BASIC METHOD)
Calculation formula .....Calculation basis of basic method
ROBERT RIELE GmbH & Co KG Page 22 5 - CALCULATION PROCEDURES