EasyManua.ls Logo

Riele 5010 V5+ - Calculation Procedure 4 (C;F;Sbrb)

Riele 5010 V5+
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5010e_513.doc / 22.01.10
5.4.4 Calculation procedure 4 (C/F/SbRb)
Method at which the difference of reagent blank A
RB
and blank of reagent blank A
RBB
regarding the amount is
subtracted of the difference of sample value A
S
and sample blank A
SB
regarding the amount, and this difference
is multiplied with a given factor F.
The sample blank A
SB
is measured before every test. The reagent blank A
RB
is entered or measured once.
Calculation procedure .......................................................CP 4
Characteristic ........................................................C / F / SbRb
Method ....................................................End Point with Factor
Calculation formula ..............C = F ( |A
S
- A
SB
| - |A
RB
- A
RBB
| )
Factor............................................................... given / entering
Reagent blank........................................entering or measuring
PHOTOMETER 5010 #7000
V5.13a 02/09/09 D
LAB.: RIELE GMBH+CO KG
DATE: 09/08/09
TIME: 10:11:15
METHOD 24: Fe
PROGRAM: 4
FACTOR: 1330
WAVELENGTH: 578nm
TEMPERATURE: 37C
MEAS. VOLUME: 900ul
WASH VOLUME: 1000ul
DELAY: 5s
MIN.UNITS: 37
MAX.UNITS: 158
UNIT: ug/dl
- - - - - - - - - - - -
MEASURE BLANK
Rb[A]: 0.085
Rbb[A]: 0.198
DELTA Rb: 0.113
NO. ABS. RESULT
1 0.715 154
Sb[A]: 0.486
- - - - - - - - - - - -
2 0.646 49
Sb[A]: 0.497
- - - - - - - - - - - -
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
ÆInsert / measure zero solution
ÆInsert / measure blank of reagent blank
ÆInsert / measure reagent blank
(Resulting blank)
ÆInsert / measure sample blank
ÆInsert / measure sample
ÆInsert / measure sample blank
ÆInsert / measure sample
ROBERT RIELE GmbH & Co KG Page 27 5 - CALCULATION PROCEDURES

Table of Contents

Related product manuals