7.2 DESCRIPTION OF UTILITY PROGRAMS
7.2.1 Dark adjustment
The dark adjustment should be done not
before the warm-up time of 15 minutes has
passed, better after one hour operation.
To avoid strange light the lid has to be
closed during the dark adjustment.
Start the dark adjustment by [START].
Automatically all eight sensitivity levels
(V0...V7) are calibrated in succession at
simultaneous display of the dark value.
The function cannot be interrupted. After com-
pletion the program returns to the utility
program level.
Monthly executed the dark adjustment
compensates possible deviations of the
measuring accuracy due to environmental
influences.
DARK ADJUSTMENT
EXIT START
DARK ADJUSTMENT
V 1 18235
PLEASE WAIT