Users Guide for DG10× 2 Series
3-20
Example for Channel Coupling
Choose CH1 as the base channel, set the phase deviation to be 10°, view the
waveforms before and after the coupling.
The operation steps:
1. Set the parameters of CH1 as shown in the following figure: Sine wave, frequency
1 kHz, amplitude 5V
PP
, Offset 0 V
DC
, phase .
Figure 3-18
Parameters Setting in CH1
2. Set the parameters of CH2 as shown in the following figure: Ramp wave,
frequency 1.5 kHz, amplitude 5V
PP
, Offset 0 V
DC
, phase .
Figure 3-19
Parameters Setting in CH2
3. Press Utility Coupling Switch Coupling On. Open the coupling function.
4. Press BaseCH Channel_1. Choose CH1 as the base channel.
5. Press PhaDev, use the keypad to input “10” and choose the unit of “°”. Set the
phase deviation to be 10°.
6. Press to save the exit the coupling setting. Press View, in the Graph Mode,
you will see the phase of CH2 has been changed to 10°.