d. Calculate the relative error of each vertical scale: |(Vavg1 - Vavg2) - (Vout1 -
Vout2)|/Full Scale x 100%.
10. Turn off CH1. Test the relative error of each scale of CH2, CH3, and CH4
respectively using the method above and record the test results.
5.4.4 Test Record From
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculatio
n Result
[1]
CH1
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
CH2
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
CH3
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
Performance Verification Test
DHO1000 Performance Verification
18
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Other manuals for Rigol DHO1204

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DHO1204 and is the answer not in the manual?

Rigol DHO1204 Specifications

General IconGeneral
BrandRigol
ModelDHO1204
CategoryTest Equipment
LanguageEnglish

Related product manuals