d. Calculate the relative error of each vertical scale: |(Vavg1 - Vavg2) - (Vout1 -
Vout2)|/Full Scale x 100%.
10. Turn off CH1. Test the relative error of each scale of CH2, CH3, and CH4
respectively using the method above and record the test results.
5.4.4 Test Record From
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculatio
n Result
[1]
CH1
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
CH2
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
CH3
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
Performance Verification Test
DHO1000 Performance Verification
18