°360×=
sourceA
RR
RR
Period
BDelayA
BPhaseA
Wherein,
PhaseA
R
B
R
represents Phase(r-r),
DelayA
R
B
R
represents Delay(r-r), and
Period
sourceA
represents the period of Source A.
6. Phase(f-f): indicates the phase deviation between the threshold middle values of
the falling edge of Source A and that of Source B. The phase formula is as follows:
°360×=
sourceA
FF
FF
Period
BDelayA
BPhaseA
Wherein,
PhaseA
F
B
F
represents Phase (f-f),
DelayA
F
B
F
represents Delay(f-f), and
Period
sourceA
represents the period of Source A.
7. Phase(r-f): indicates the phase deviation between the threshold middle values of
the rising edge of Source A and the falling edge of Source B. The phase formula is
as follows:
°360×=
sourceA
FR
FR
Period
BDelayA
BPhaseA
Wherein,
PhaseA
R
B
F
represents Phase (r-f),
DelayA
R
B
F
represents Delay(r-f), and
Period
sourceA
represents the period of Source A.
8. Phase(f-r): indicates the phase deviation between the threshold middle values of
the falling edge of Source A and the rising edge of Source B. The phase formula is
as follows:
°360×=
sourceA
RF
RF
Period
BDelayA
BPhaseA
Wherein,
PhaseA
F
B
R
represents Phase (f-r),
DelayA
F
B
R
represents Delay(f-r), and
Period
sourceA
represents the period of Source A.
TIP
• Source A and Source B can be any channel among CH1-CH4, D0-D15, and Math1-
Math4.
• The default threshold middle value is 50%.
Measurements
DHO900 User Guide
138
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30

Table of Contents

Other manuals for Rigol DHO924

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DHO924 and is the answer not in the manual?

Rigol DHO924 Specifications

General IconGeneral
BrandRigol
ModelDHO924
CategoryTest Equipment
LanguageEnglish

Related product manuals