EasyManuals Logo

Rigol DHO924 User Manual

Rigol DHO924
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #193 background imageLoading...
Page #193 background image
Set the Modulation Frequency
When the selected modulating waveform is not noise, you can set the modulation
frequency. Click or tap the Mod Freq input field to set the modulation frequency with
the pop-up numeric keypad. You can also use the corresponding multipurpose knob
to set the value. Its range is from 2 mHz to 1 MHz.
To Enable the Modulation Output
Click or tap Modulation on/off switch to enable the modulation function. In the
"Basic waves" setting menu, click or tap the Output on/off switch to enable the
waveform output. The instrument will output the modulated waveform via the rear-
panel [AFG OUT] connector according to the settings.
Function/Arbitrary Waveform Generator (for
DHO914S and DHO924S only)
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO900 User Guide
177
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30

Table of Contents

Other manuals for Rigol DHO924

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DHO924 and is the answer not in the manual?

Rigol DHO924 Specifications

General IconGeneral
BrandRigol
ModelDHO924
CategoryTest Equipment
LanguageEnglish

Related product manuals