38Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
BUMP.SET (page 56)
(cont’d)
BP.EXPD
(page 60)
Defines what action must be taken if a bump test is due upon
startup.
CONFIRM
(factory setting): Press and release AIR to acknowl-
edge that bump test is due and continue to Measuring Mode.
CANT.USE
: Cannot enter Measuring Mode until a successful
bump test is performed.
NONE
: A screen indicates that bump test is due but warmup
sequence continues.
ESCAPE Return to the BUMP SET item in User Mode.
ALARM-P (page 61) Set alarm points (WARNING, ALARM, ALARM H, STEL, TWA) and/or reset all
alarms to their default settings.
LUNCH (page 63) ON
: Lunch break feature is on. Instrument asks if you want to resume TWA and
PEAK readings at startup.
OFF
(factory setting): Lunch break feature is off. Instrument resets TWA and
PEAK readings every time it’s turned on.
BEEP (page 63) Set confirmation beep parameters.
BEEP.SEL
(page 64)
LED
: LED flashes and instrument vibrates based on interval
defined in BEEP.INT to confirm instrument is still operating.
BUZZER
: Buzzer sounds and instrument vibrates based on inter-
val defined in BEEP.INT to confirm instrument is still operating.
LED+BUZ
: LED flashes, buzzer sounds, and instrument vibrates
based on interval defined in BEEP.INT to confirm instrument is
still operating.
BMP/CAL
: LED flashes based on interval defined in BEEP.INT
if bump test or calibration is due regardless of whether BP.RMDR
and/or CAL.RMDR are set to ON.
OFF
(factory setting): No alerts to confirm instrument is still oper-
ating or that a bump test or calibration is due.
BEEP.INT
(page 64)
Confirmation alert interval. Confirmation type defined in
BEEP.SEL.
Options: 0.5 minute and 1 to 99 minutes in 1 minute increments.
The factory setting is 5 minutes.
ESCAPE Return to the BEEP item in User Mode.
BL TIME (page 65) How long the back light stays on after the last button press.
Options: 0 - 255 seconds or OFF. The factory setting is 30 seconds.
KEY.TONE (page 65) ON
(factory setting): Buzzer sounds when button is pressed.
OFF
: Buzzer does not sound when button is pressed.
Table 8: User Mode Menu Items