60Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
BP.EXPD
This item defines what indicator is given during start up when a bump test is due and BP.RMDR
is set to ON (factory setting is OFF).
CONFIRM (factory setting): The HS-04 gives an indicator at start up if a bump test is past due.
Press and release AIR to continue without bump testing or press and release POWER MODE to
enter User Mode and perform a bump test.
CANT.USE: The HS-04 gives an indicator at start up that a bump test is past due. Press and
release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no effect.
A successful bump test must be performed in order to use the instrument.
NONE: The HS-04 gives an indicator at startup that a bump test is past due. It is not necessary to
acknowledge the indicator. If desired, press POWER MODE to enter User Mode and perform a
bump test. The warm-up sequence continues on its own.
1. After entering the BUMP.SET menu, press AIR to scroll to BP.EXPD.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.EXPD item in the
BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.
Exiting the BUMP.SET Menu
1. While in the BUMP.SET menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the BUMP.SET item in User Mode.
3. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
BP .EXPD