HS-04 Operator’s Manual Chapter 5: User Mode and Calibration • 59
BP.RMDR
ON: The HS-04 gives an indicator at start up if it is due for bump testing. The type of indicator
depends on the BP.EXPD setting (see page 60). If the instrument is not due for bump testing, a
check mark appears in the lower left corner of the LCD.
OFF (factory setting): The HS-04 does not give an indicator at start up if it is due for bump
testing.
1. After entering the BUMP.SET menu, press AIR to scroll to BP.RMDR.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.RMDR item in the
BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.
BP.INT
This setting defines the amount of time between bump tests and can be set in 1 day increments.
The minimum setting is 0 days and the maximum setting is 30 days (factory setting).
1. After entering the BUMP.SET menu, press AIR to scroll to BP.INT.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.INT item in the
BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.