58 • Chapter 5: User Mode and Calibration HS-04 Operator’s Manual
CAL.TIME
The CAL.TIME is the total time the instrument is exposed to calibration gas if A-CAL is set to
ON and a bump test fails. The bump test time is deducted from the calibration time. For example,
if the CAL.TIME is set to 90 seconds and the GAS.TIME is set to 30 seconds, if the bump test
fails, the HS-04 is only exposed to gas for an additional 60 seconds. The available values are 90
seconds (factory setting), and 120 seconds.
1. After entering the SETTING menu, press AIR to scroll to CAL.TIME.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.TIME item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 58 to return to BUMP.SET menu.
A-CAL
ON (factory setting): If a bump test fails, the unit automatically begins a calibration.
OFF: If a bump test fails, the unit does not automatically begin a calibration.
1. After entering the SETTING menu, press AIR to scroll to A-CAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the A-CAL item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 58 to return to BUMP.SET menu.
Exiting the SETTING Menu
1. While in the SETTING menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the SETTING item in the BUMP.SET menu.
3. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.