EasyManua.ls Logo

ROCAMED SIRIUS - 4.6 Installation Procedure

ROCAMED SIRIUS
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
DGM001519.01 UM Sirius CE/ROTW
Installation procedure
The installation procedure must be performed each time the device is installed for the first time or after being
transported by means of cars, elevators, trucks, air cargo, etc.
During installation the device must be checked for proper operation and possible malfunctions after
transportation of the laser device must be corrected.
The installation procedure includes also a training course from the distributor to the user concerning the use of
the medical device.
The first procedure step typically takes several hours, during this time the access to the installation site is
forbidden. The case is normally shipped to the distributor.
It is extremely important that the packed materials are checked immediately upon their arrival, if possible, in
the presence of the ƐŚŝƉƉŝŶŐĂŐĞŶƚƐ employee, as follows:
x Open the packaging and put the laser device in a proper site for a general check
x Execute the following operations for the general check:
o Check the labels of the device
o Connect the remote door interlock
o Connect the footswitch
o Connect the laser device to the power supply
o Turn on the system
o Check the system and verify if alert messages are displayed
o Connect the RFID fiber and wait that the Sirius system recognizes the type and the number of
uses of connected fiber
o Check the system and verify if alert messages are displayed
o Change the status of the Sirius laser system in Ready
o Check the system and verify if alert messages are displayed
o Change the status of the Sirius laser system in Standby
o Turn off the system
x After the general check:
o Remove the optical fiber
o Remove the footswitch
o Remove the interlock
o Remove the key
Note: Quanta System advises wrapping the device with a large quantity of protective plastics.
Note: The shipment of the device to the final destination of the customer is under the responsibility of the
distributor. Quanta System is not responsible for any possible damage caused during this phase.
x Install the device in the room indicated by the customer in the following way:
o Place the device so that the operator can have easy access to operate on the mains cable by
connecting/disconnecting it when necessary
o Connect the device to the power supply
o Connect the interlock connector
o Connect the footswitch
o Check the laser device
x Perform further controls or additional tests.
x Perform a training to the end user on the following items:
o Proper fiber attachment
o Operation of the device
Caution: Do not start any action with the laser device before the official personnel have performed the
installation procedure. The warranty does not cover any damage to the laser device before the installation.