DGM001519.01 UM Sirius CE/ROTW
D/EdEE>E/E'E^dZ/>/d/KE
This section contains mainly user information regarding the care and cleaning of the device. Should you require
any assistance concerning maintenance or technical intervention, please contact Quanta System (or refer to the
Service Manual to find more information).
The laser, cooling system and control electronics are enclosed in a tamper-resistant console.
Caution: The laser instrument does not contain any component that the user is authorized to modify or
repair. Only service personnel authorized by the manufacturer can perform any modification of the device
including the replacement of the power supply cable.
Device cleaning
The Sirius laser device does not need special maintenance by the user. Switch off the device before starting the
cleaning procedure. Clean the visible surfaces with a damp cloth, taking care not to let the water enter the
device (in particular attention must be paid in not spilling any drop on monitor frame interstice and monitor
itself). Do not use alcohol or disinfectant solutions, because they are highly flammable. While cleaning, be careful
not to let the cleaning solution leak in the fiber connection port. Use the supplied caps to close the fiber
connection port after each use. Do not use any alcohol solution to clean the display.
Laser Maintenance and technical check
The Sirius Laser device is designed for maximum safety and performance. Under normal operating conditions
and when respecting a careful use, the manufacturer recommends a check-up of the device by a qualified
technician every 12 months.
The intensive use, dust, or continuous movement of the laser in different places may require a more frequent
monitoring.
Safety labelling check
The user must regularly verify the integrity and readability of the security labels placed on the device.
If labels are damaged, they must be replaced immediately in accordance with the plan described
in paragraph 2.13, Labelling plan.
Check of the line cable
The Sirius device has a detachable line cable. The line cable can be subject to deterioration over time and
therefore must be checked periodically its status.
Warning: Only the Manufacturer or trained and qualified personnel are permitted to substitute the line
cable, in case of deterioration.
Optical fiber maintenance
Warning: Be sure that fiber sterilization is not expired (expiration date is reported on fiber label).
Please read carefully the instruction manual of the optical fiber before use to assure a proper and safe use,
maintenance and reprocessing, if applicable.