Shutdown procedure and protection against unauthorized use
Once device operation is accomplished and the laser is in Standby mode, one can proceed with its shutdown
procedure as follows:
x Disconnect the optical fiber from the device and cover the laser device output connector with its
dedicated protection shutter
x Press the frontal ON/OFF button
x Turn the key switch to the configuration and remove the key to prevent unauthorized uses
x Turn off the main switch on the rear panel and unplug the power cable
x Disconnect the remote interlock
x Disconnect the footswitch
x In order to disconnect the device from the mains it is necessary to detach the supply cable plug from the
mains.
x Keep the device and its accessories in a dry and safe place
In order to avoid improper use of the device, keys shall be removed when the device is not being used.
Mains disconnection
In order to disconnect the device from the mains it is necessary to detach the supply cable plug from the mains.