While delivering radiation, values of emitted Energy, in Joules, (A) and Lasing Time (B), will be increased. To
reset the pulses/energy counters, refer to Section 5.2.1.5.
At the end of the treatment, release the footswitch and enter the Standby mode by pressing the Ready/Standby
area on the display, or the dedicated button on the footswitch.
Note: If footswitch pedal is kept released for a long time during the READY mode, the system will automatically
enter the STANDBY mode.
Warning: While keeping the fiber tip inside of patient body, user should switch from Ready to
Standby status when performing provisionally any procedure other than lasing.
Warning: Before removing the fiber from patient body, user MUST switch from Ready to Standby
status (both when treatment is completed and when performing provisionally any procedure other
than lasing).
Laser Parameters
A wide range of combination of pulse energy (Joules, J) and frequency (Hertz, Hz) is available. The laser output
power (Watt, W) for each possible combination is calculated (max 60 W) as follows:
Power (W) = Energy (J) × Frequency (Hz)
The diameter, type and number of uses of the connected fibers are automatically detected. Optical fibers with
ĚŝĂŵĞƚĞƌчϮ00 ʅŵŵĂLJŚĂǀĞůŝmitations in the output power settings (energy / frequency) due to their technical
features.
Error and warning description
Different errors may be displayed on the control panel. Every time an error occurs, a message appears at the
top of the screen specifying the error type:
When an error occurs, the System behaviour is different depending on the issue level of severity:
ERROR Status: /ŶƚŚĞZZKZƐƚĂƚƵƐƉŽǁĞƌĞůĞĐƚƌŽŶŝĐƐĂƌĞĐƵƚŽĨĨƚŚƌŽƵŐŚĂƐƵŝƚĂďůĞƐǁŝƚĐŚĂŶĚƚŚĞĚĞǀŝĐĞ
starts running in a safe mode or gets restarted. This Error occurs every time the Control finds the System in a