EasyManua.ls Logo

ROCAMED SIRIUS - Page 65

ROCAMED SIRIUS
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65
DGM001519.01 UM Sirius CE/ROTW
IEC 60601-1-2 Table 202
'ƵŝĚĂŶĐĞĂŶĚŵĂŶƵĨĂĐƚƵƌĞƌƐĚĞĐůĂƌĂƚŝŽŶʹ electromagnetic immunity
The equipment SIRIUS is intended for use in the electromagnetic environment specified below. The
customer or the end user of the SIRIUS should assure that it is used in such an environment
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 KV contact
±2, 4, 8, 15 KV air
Compliant
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 KV for power supply
lines
±1 KV for I/O lines
Compliant
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment
Surge
IEC 61000-4-5
±0.5, 1 KV for power
supply lines (line to
line)
±0.5, 1, 2 KV for power
supply lines (line to
earth)
±2 KV for I/O lines
Compliant
Compliant
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment
Voltage Dips, Short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Dips:
0% Ut for 0,5 cycle
0% Ut for 1 cycle
70% Ut for 25 cycles
(50 Hz)
70% Ut for 30 cycles
(60 Hz)
Interruptions:
0% Ut for 250 cycles
(50 Hz)
0% Ut for 300 cycles
(60 Hz)
Compliant
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of the SIRIUS
requires continued operation
during power mains
interruptions, it is
recommended that the SIRIUS
be powered from an
Uninterruptible Power Supply
or Battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
Compliant
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment
Note: Ut is the AC mains voltage prior to application of the test level

Table of Contents