Pos No Field Type Max EV CV
Comments
Elecsys Type Format
cobas Type Format
4 (4) CM 30/20 X R Indicates ordered sample.
Setting is as follows:
<SequenceNo>^<CarrierNo>^
<PositionNo>^^<SampleType>^
<ContainerType>
<SequenceNo> Type: NM Max: 4
indicates cobas e 411 analyzer.
internal sequence No.
<CarrierNo> Type: ST Max: 5
indicates carrier No. (Disk/Rack.)
<PositionNo> Type: NM Max: 2
indicates position No. in carrier.
<SampleType> Type: ST Max: 7
indicates sample type.
SAMPLE: patient sample.
CONTROL: control sample.
<ContainerType> Type: ST Max: 7
indicates type of sample cup.
NORMAL: test tube or sample
cup.
REDUCED: sample cup, only.
Indicates ordered sample.
Setting is as follows:
<SequenceNo>^<CarrierNo>^
<PositionNo>^^<SampleType>^
<ContainerType>
<SequenceNo> Type: NM Max: 4
indicates cobas e 411 analyzer.
internal sequence No.
<CarrierNo> Type: ST Max: 5
indicates carrier No. (Disk/Rack.)
<PositionNo> Type: NM Max: 2
indicates position No. in carrier.
<SampleType> Type: ST Max: 2
(*2) indicates sample type
S1: blood serum.
S2: urine.
S5: others.
QC: control sample.
<ContainerType> Type: ST-Max: 7
(*1) indicates type of sample cup.
SC: test tube or sample cup.
MC: reduced sample volume.
5 (5) Universal Test ID CM 8 X X Indicates order.
Repeats up to 60 orders. Only up to
18 for test selection information are
avaiölable. Setting is as follows:
^^^<ApplicationCode>^
<Dilution>\
Repeat delimiter \ for multiple test
selection. For upload, ALL is fixed.
<ApplicationCode> Type: NM
Max: 3 (*4), (*6) indicates e411
Application No.
<Dilution> Type: NM Max: 1
indicates automatic dilution factor.
(*5)
0: ratio=1 (not diluted).
1: ratio=2
2: ratio=5
3: ratio=10
5: ratio =20
7: ratio=50
9: ratio=100
Indicates order.
Repeats up to 60 orders. Only up to
18 for test selection information are
available. Setting is as follows:
^^^<ApplicationCode>^
<Dilution>\
Repeat delimiter \ for multiple test
selection.
<ApplicationCode> Type: NM
Max: 3 (*4), (*6) indicates e411
host test No.
The host test no. can be a test
code or a calculated test no (only
for upload).
<Dilution> Type: NM Max: 1 (*5)
indicates automatic dilution ratio.
(also for no dilution attach "^" after
Host Test No. like ^^^10^)
None or
1: ratio=1 (not diluted)
2: ratio=2
5: ratio=5
10: ratio=10
20: ratio=20
50: ratio=50
100: ratio=100
6 (6) Priority ST 1 X R
Indicates priority of patient
samples
R: routine sample.
S: stat sample
Indicates priority of patient
samples
Not used for control samples.
R: routine sample.
S: stat sample
7 Requested/Ordered
Date and Time
8 (7) Specimen Collection
Date and TIme
TS 14 X X Indicates reception date and time of request. Setting is as
follows:
Optional
YYYYMMDDHHMMSS
9 Collection End Time
10 Collection Volume
11 Collector ID