Page #50 background imageLoading...
Page #50 background image
cobas e 411 analyzer
Roche Diagnostics
50
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Test selection information in real time
An example of communication trace is shown below.
The following trace is shown without Start ([STX][FN]), End
([CR][EXT][CS1[CS2][CR][LF]], and response from each receiver
([ENQ][ACK][EOT]).
The HOST is required to send back the SequenceNo, CarrierNo, and Position to the
cobas e 411 analyzer without changing any of them.
If any information is changed, the analyzer sends a cancelation message to the HOST
and the measurement is skipped.
Disk Type
The
cobas e
411 analyzer sends test selection information such as sample
ID=000004, sequence No.=40, Disk No=0, and Position=5.
H|\^&||||||||||P||[CR]
Q|1|^000004^40^0^5^^SAMPLE^NORMAL||ALL||||||||O[CR]
L|1|I [CR]
Example 1: HOST replies test selection information of sample ID=000004.
H|\^&||||||||||P||[CR]
P|1 [CR]
O|1|000004|40^0^5^^SAMPLE^NORMAL|^^^10^\^^^30^2\^^^40^|R||||||N
||||||||||||||Q
L|1| [CR]
Example 2: HOST replies without order information of sample ID=000004.
H|\^&||||||||||P||[CR]
P|1 [CR]
O|1|000004|40^0^5^^SAMPLE^NORMAL||R||||||N||||||||||||||Z[CR]
L|1| [CR]
The
cobas e
411 analyzer sends a cancel to HOST when HOST does not reply
within 15 seconds.
H|\^&||||||||||P||[CR]
Q|1|^000004^40^0^5^^SAMPLE^NORMAL||ALL||||||||A[CR]
L|1|I [CR]