EasyManua.ls Logo

Roche cobas e 411 - Result Value Is a Qualitative Test; Control Sample

Roche cobas e 411
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
cobas e 411 analyzer
Roche Diagnostics
54
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Result value is a qualitative test
The cobas e 411 analyzer sends a test result of sample ID=000010, sequence
No.=442, Rack No.=0005, and Position=1.
H|\^&||||||||||P||[CR]
P|1|||||||||||||||||||||||||||||||||| [CR]
O|1|000010|442^0005^1^^SAMPLE^NORMAL|ALL|R|20051220104418|||||X
||||||||||||||O[CR]
R|1|^^^400^^0|-1^0.303|COI|^|N||F|||20051220110334|
20051220112404|[CR]
L|1| [CR]
The cobas e 411 analyzer sends a test result of sample ID=000010, sequence
No.=442, Rack No.=0005, and Position=1.
H|\^&|||cobas-e411^1|||||host|RSUPL^REAL|P|1[CR]
P|1 [CR]
O|1|000010|442^0005^1^^S1^SC|^^^400^|R||||||N||||1|||||||
20051220104418|||F[CR]
R|1|^^^400//not|-1^0.303|COI||N||F||admin|||E1[CR]
L|1|N[CR]
Control sample
H|\^&||||||||||P||[CR]
P|1|||||||||||||||||||||||||||||||||| [CR]
O|1|PC U2|96^0019^1^^CONTROL^NORMAL|ALL|R|20051220104418|||||X\Q
||||||||||||||O[CR]
R|1|^^^10^^0|1.45|ulU/ml|1.37^1.97|N||F|||20051220110334|
20051220112404|[CR]
L|1| [CR]
H|\^&|||cobas-e411^1|||||host|RSUPL^REAL|P|1[CR]
P|1 [CR]
O|1|PC U2|96^0019^1^^QC^SC|^^^400^|||||||Q||||1
|||||||20051220104418|||F[CR]
R|1|^^^400//not|1.26^|ulU/ml||L||F||admin|||E1[CR]
L|1|N[CR]

Related product manuals