cobas e 411 analyzer
Roche Diagnostics
52
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Rack Type
The cobas e 411 analyzer sends inquiry for sample ID=000002, sequence No.=3
Rack No.=@95 and Position=2.
H|\^&||||||||||P||[CR]
Q|1|^000002^3^@95^2^^SAMPLE^NORMAL||ALL||||||||O[CR]
L|1|I[CR]
HOST replies test selection information of sample ID=000002.
H|\^&||||||||||P||[CR]
P|1 [CR]
O|1|000002|3^@95^2^^SAMPLE^NORMAL|^^^10^|R||||||N||||||||||||||Q[CR]
L|1| [CR]
The cobas e 411 analyzer sends inquiry for sequence No.=3 Rack No.=0007,
Position=2, and no sample ID.
H|\^&||||||||||P||[CR]
Q|1|^^3^0007^2^^SAMPLE^NORMAL||ALL||||||||O[CR]
L|1|I[CR]
Key information for a non-barcoded sample is Rack No and position. If barcode read error takes
place for a barcoded sample, the key information is also Rack No and position. In such cases,
the cobas e 411 analyzer inquires after reading Rack No.
The
cobas e
411 analyzer sends inquiry for sample ID=000002, sequence No.=3
Rack No.=@95 and Position=2.
H|\^&|||cobas-e411^1|||||host|TSREQ^REAL|P|1[CR]
Q|1|^^000002^3^@95^2^^S1^SC||ALL||||||||O[CR]
L|1|N[CR]
HOST replies test selection information of sample ID=000002.
H|\^&|||host^1|||||cobas-e411|TSDWN^REPLY|P|1[CR]
P|1 [CR]
O|1|000002|3^@95^2^^S1^SC|^^^10^|R||||||A||||1||||||||||O[CR]
L|1|N[CR]
The cobas e 411 analyzer sends inquiry for sequence No.=3 Rack No.=0007,
Position=2 and a generated sample id=@3.
H|\^&|||cobas-e411^1|||||host|TSREQ^REAL|P|1[CR]
Q|1|^^@3^3^0007^2^^S1^SC||ALL||||||||O[CR]
L|1|N[CR]
When it is rack type without sample ID, incl. barcode read error, the cobas e 411 analyzer
inquires after reading Rack No.
If a barcode read error occurs and the rack no. is generated, the sample cannot be identified by
the host.
When the rack type communication for a rack does not complete within approximately
40 seconds, the cobas e 411 analyzer sends a cancel to HOST.