cobas e 411 analyzer
Roche Diagnostics
28
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Record structure
The
cobas e
411 analyzer host protocol is compatible with the
cobas®
6000
analyzer series protocol with some minor differences.
No. Object Remarks
1
Carrier no.
cobas e 411 rack analyzer:
When rack no. is unknown, @ is attached in front of <CarrierNo> followed by a delivered value.
cobas e 411 disk analyzer:
The carrier no. is always available on a disk system.
2
Sample ID
When barcode read error occurs, the sample ID is generated as follows: @<Sequence no.>
If a sample barcode read error occurs on a rack system and the rack no. is generated, the sample
cannot be identified by the host.
3 Container Type MC is not micro cup, but indicates reduced value.
4 Result Record cobas type format transmits calculated tests. The host test code numbers can be defined.
5
Comment Record (following
the order record)
The comment record, that follows the order record for patient demographic data is not used.
6
Comment Record (following
the result record)
When no data alarm for message value is available, no comment record is transmitted to host.
Table 7: Record structure