of 61 Host Interface Manual for use in the US
Manual version Revision date Changes
2.1 April 2011 Additional screenshots and remarks on page 5
Host Test Code Setting.
2.2 November 2013 • Updated screenshots, on page 5 Host
Communication Setting.
• Additional explanation, on page 8 STAT
Sample Result Upload Setting.
• Note (*6) no sending of multiple orders from
the same test, on page 37 ff – valid from
Software version 02-04.
2.3 March 2016 Modified tip for test selection in real time in the
Communication trace chapter.
• Used global document version 2.3 as
reference.
• Corrected branding, as necessary.
• Corrected English, as necessary.
This document is for users of the cobas e 411 analyzer.
The contents of this document, including all graphics and photographs, are the
property of Roche. No part of this document may be reproduced or transmitted
in any form or by any means, electronic or mechanical, for any purpose, without
the express written permission of Roche.
Every effort has been made to ensure that the information is correct at the time
of publishing. Not all functionality described in this manual may be available to
all users. Roche Diagnostics reserves the right to change this publication as
necessary and without notice as part of ongoing product development. Such
changes may not immediately be reflected in this document.
This document is intended for the US market only.
For prescription use only.
This document describes communication procedure related to communication
method that enables intercommunication between Laboratory Host System,
hereinafter referred to as HOST, and the cobas e 411 analyzer. Specification and
software described herein comply with the following ASTM communication
protocol (HOST communication ASTM higher-layer: High-Level, lower-layer:
Low-Level I/F specification): Specification X12 of ASTM (American Society of
Testing and Materials).
ASTM E1381-91: Specification for Low-Level Protocol to Transfer Messages
Between Clinical Laboratory Instruments and Computer
System
ASTM E1394-91: Standard Specification for Transferring Information
between Clinical Instruments and Computer System
As cobas e 411 analyzer is a successor version of the Elecsys 2010 analyzer, it
upholds Elecsys 2010 analyzer HOST communication protocol. Further, a new
communication protocol is added to keep compatibility with cobas® analyzers.
These two (2) communication protocols such as Elecsys type and cobas type are
selectable. MSRs (Manufacturer Specific Records) that are the original protocol
of Elecsys 2010 analyzer are not supported by cobas e 411 analyzer.
In case of communication problems between a host and the cobas e 411
analyzer please copy the trace file d:\e411\data\log\host_trace.log onto a
removable media and contact the Customer Support Center at 1-800-428-2336.