cobas e 411 analyzer
Roche Diagnostics
42
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Result record
Elecsys type (upload)
R|1|^^^10^^0|0.310|ulU/ml|0.270^4.20|N||F|||20050619094203|20050619101521[CR]
(1)(2) (3) (4) (5) (6) (7) (8) (10) (11)
cobas type (upload)
R|1|^^^10//not|0.310|ulU/ml||N||F||admin|||E1[CR]
(1)(2) (3) (4) (5) (7) (8) (9)
Pos No Field Type Max EV CV
Comments
Elecsys Type Format
cobas Type Format
1 (1) Record Type ID ST 1 X X R is fixed.
2 (2) Sequence Number NM 6 X X Indicates sequence No.
3 (3) Universal Test ID CM 10/20 X X Indicates order.
^^^<ApplicationCode>^
<Dilution>^<pre-dilution>
<ApplicationCode> Type: NM
Max: 3 indicates e411 Application
No.
<Dilution> Type: NM Max: 1
indicates automatic dilution factor
when ordering.
< pre-dilution > Type: ST Max: 1
0: without pre-dilution.
1: with pre-dilution.
Indicates order.
^^^<ApplicationCode>/<Dilution
>/<pre-dilution>
<ApplicationCode> Type: NM
Max: 3 indicates e411 host test no.
The host test no. can be a test
code or a calculated test no.
Calculated tests can be defined on
the Calc. Test screen. A total of 5
calculated tests can be set. When
all tests requiring calculation are
ordered and successfully
measured, the analyzer
automatically transmits the
calculated tests.
However, the host may not send a
calculation test order.
<Dilution> Type: ST Max: 1
indicates automatic dilution factor
when ordering: 1, 2, 5, 10, 20, 50,
100
< pre-dilution > Type: ST Max: 11
not: without pre-dilution.
pre-diluted: with pre-dilution.